BOD Incubator (Cooling)
Specification
- Temperature range: 5.0°C to 60.0°C. Accuracy / Uniformity: ±0.5°C/±1°C
- Temperature display: Digital LCD & Safety Controlller display LED 3½ digit
- Control: Microprocessor based PID control. With Auto tune
- Resolution: 0.1°C
- Temperature sensor: PT 100
- Heating: ‘U’ shaped Nichrome wire heater in SS sheathing
- Cooling: CFC free Copeland make compressor utilizing R134A eco friendly refrigerant, with condenser,motor and relay.
- Air circulation: Flange motor with impeller / blower
- Construction: Double wall, with insulation, outer door key lockable, inner glass viewing door
- Chamber illumination: Fluorescent light with door switch
- Safety Instructions
- High temperature safety cut off
- Electrical overload cut off relay for compressor
- Time delay for compressor switch ON
- Electrical circuit breaker
- Feet: Castor wheels with leveling screws
- Trays: SS wire mesh heavy duty
- Electrical: 230V / 15A / 50 Hz
- Calibration certificate by ERTL – Mumbai with traceability to NPL – New Delhi
Dimentions
Inner Chamber Dimesion |
External Dimension |
Capacity |
Wattage |
Shelves |
W x D x H cm |
W x D x H cm |
Liters |
|
(SS) |
35 x 35 x 35 |
55 x 60 x 85 |
72 |
1500 |
2 |
45 x 45 x 45 |
65 x 70 x 95 |
90 |
2000 |
2 |
45 x 45 x 60 |
70 x 70 x 105 |
120 |
2000 |
3 |
60 x 60 x 60 |
90 x 85 x 110 |
215 |
2500 |
3 |
60 x 60 x 90 |
90 x 85 x 135 |
320 |
3000 |
3 |
60 x 60 x 125 |
90 x 85 x 170 |
450 |
3500 |
4 |
Optional Extras / Accessories
- PLC based system: PLC with HMI 3.7” touch screen grey monitor with 21 CFR part 11 features software
- Magnetic door lock feature
- GSM system: alarms sent to predefined mobile numbers
- HMI : 5.7” touch screen grey in lieu of 3.7”
- Digital PID controller with printer interface to connect EPSON Dot matrix line printer. Print interval programmable, can print date, time, temperature.
- PC interface for above controller with software on CD & hardware or software with 21 CFR part 11
features
- Imported RH sensor European make, Rotronic Switzerland for higher accuracy
- Standby humidity system (except for 52 and 120 liters)
- Standby refrigeration system (except for 52 and 120 liters)
- Data logger: 4 point temperature data logger with sensors placed at specific points in chamber with printer interface and memory of 6000 readings per channel.
- PC interface for above controller with software on CD & hardware
STD MODEL |
GMP MODEL |
1) Outer cabinet mild steel powder coated. |
1) Inner chamber stainless steel 304 grade. |
2) Inner chamber stainless steel SS 304 grade. |
2) Inner chamber stainless steel mirror finish |
We provide DQ, IQ, OQ, PQ documentation. |
Pre Installation Mandatory Requirement :
(Guarantee void in case of non compliance)
- Room temperature around machine preferably at 25°C with air conditioning or a well ventilated room with exhaust fan. However surrounding temperature should not exceed 30°C
- Stabilized input voltage of 230V AC 20 Amps. Use of servo controlled stabilizer is recommended
|